Technical specs
Measuring | Continuously CO in the gaseous streams. |
Service | Safety in spray drying food and feed, and other drying processes. |
Sampling system | Atmospheric. Sample conditioning with condense coolers. |
Measuring principle | NDIR gas filter correlation |
Sample probes | |
Stainless steel sintered filter probes | High-capacity filter probe, Varinline® compatible, DN 25 |
Maximum process temperature | 450ºC |
Sample line | 6mm or 8mm outside diameter (no heating required). |
Measuring ranges | |
Measuring range CO | 0 – 200 PPM. |
Lower detectable limit CO | < 0.04 PPM or 0.2% of reading whichever is greater. |
Zero drift | 30 days: < 0.1 PPM. |
Span drift | 7 days: < 0.5% of reading. |
Precision CO | 0,1 PPM or 0.1% of reading, whichever is greater. |
Sample flow | 5 normal liter/min per sample stream. |
Outputs | |
CO | PPM absolute and delta CO. |
Alarm levels CO | Three levels; free adjustable. |
Analogue outputs | Up 4x 4-20 mA (ref. CO and DCO) |
Digital outputs | Minor alarms |
Major alarms | |
CIP acknowledges | |
System ready | |
System on | |
Interface | HMI to be installed in control room or on cabinet door. |
Installation and utilities | |
Access cabinet | Front and back door. |
Weight | Approx 300 kg |
Mounting | Floor |
Hoisting | 4 lifting eyes |
Dimensions cabinet | 2000 x 600 x 820 mm; 81.1” x 23.6” x 36.4” (h x w x d) |
Ambient cabinet | max 30°C |
Power | 115-230VAC |
Power consumption | 2500W |
Sound pressure level | < 70 dB(A) |
IP Rating | IP54 |